Optimal Certified Power Analyzer Up Till Now

Ultimate Weapon for Testing and Certification Laboratory Analysis

Watch the video Learn more about parameters>

PA8000 Power Analyzer Parameters

Input parameters:

Input parametersParameter description
Measurement rangeVoltage1.5V,3V,6V,10V,15V,30V,60V,100V,150V,300V,600V,1000V
CurrentDirect input (Crest factor is 3)5A:10mA、20mA、50mA、100mA、200mA、500mA、1A、2A、5A
Sensor input (Crest factor is 3)50mV、100mV、200mV、500mV、1V、2V、5V、10V
Maximum continuous allowable input valueVoltageChoose the lower value of 2000V peak value and 1300V voltage RMS
CurrentDirect input5A:Choose the lower value of 15A peak value and 6.5A RMS
Sensor inputPeak value less than or equal to 4 times of the range; RMS value less than or equal to 2 times of the range
Maximum transient allowable input value
(≤ 1s)
VoltageChoose the lower value of 3000V peak value and 1500V RMS
CurrentDirect input5A:Choose the lower value of 22.5A peak value and 10A RMS
Sensor inputPeak value less than or equal to 5 times of the range; RMS value less than or equal to 3 times of the range
Input parameterPA8000
Input bandwidthDC,0.1Hz~5MHz
Sampling rate2MS/s
Maximum continuous common mode voltage1000Vrms

Accuracy:

Frequency range of input signalPA8000
Voltage / CurrentPower
DCCurrent:0.05+0.05
Voltage/Sensor:0.05+0.05
Current:0.05+0.10
Voltage/Sensor:0.05+0.10
0.1Hz ≤ f < 30Hz0.03+0.050.08+0.10
30Hz ≤ f < 45Hz0.03+0.030.05+0.05
45Hz ≤ f < 66Hz0.01+0.020.01+0.03
66Hz ≤ f < 1KHz0.03+0.030.05+0.05
1KHz ≤ f < 10KHz0.10+0.050.15+0.10
10KHz ≤ f < 50KHz0.20+0.100.30+0.20
50KHz ≤ f < 100KHz0.50 + 0.300.70 + 0.50
100KHz ≤ f < 500KHz(0.004f+0.8) + 0.5(0.02f-0.3) + 1.0
500KHz ≤ f ≤ 1MHz(0.01f-2.2) + 1.0(0.042f-12) +2.0

The measurement accuracy of the power analyzer is calculated under the following conditions: Temperature: 23±1℃; Humidity: 30-70% RH; Input signal: sine wave; Common mode voltage: 0V; After preheating, line filter: OFF; λ (Power factor): 1; Crest factor: 3. After preheating for 30 minutes, f = frequency. Data update rate: 500 ms.

Display:

Display ParameterPA8000
Display12.1 inches color LCD display
Resolution1280×800 pixels
Touch screenSupport for touch screen operation
Display update rateSame as the data update rate

Storage:

Internal SSD capacity 60G storage capacity supports long-time storage of more than 10,000 hours (normal)
USB storage interfaceSupport for USB storage interface

Measurement Function/Measurement Condition:

Data update rateSelect from 1ms、10ms、50ms、100ms、250ms、500ms、1s、2s、5s、10s、20s
Display update rateSame as the data update rate
Response timeSame as the data update rate

Harmonics measurement (PLL synchronous source method):

Fundamental frequency of PLL sourceSampling rate(S/s)Window width relative to FFT data length(fundamental frequency)Maximum harmonic analysis orderNumber of sampling points
0.5~1Hzf×819215008192
1~5Hzf×409625008192
5~10Hzf×204845008192
10~640Hzf×102485008192
640~1.28kHzf×512162558192
1.28kHz~2.56kHzf×256321008192
2.56kHz~5kHzf×12864508192

Motor function –analog quantity input parameters :

Input modeSafety BNC, floating, isolation, electrical isolation among A, B, Z of TORQUE and SPEED
Input impedance1MΩ±100kΩ
Range1V、2V、5V、10V、20V
Cut-off frequency (configurable)100Hz,10kHz,50kHz,OFF
Effective measurement range±110%
Maximum allowable voltage±22V
Maximum common mode voltage±42Vpeak
Number of bits16bit
Sampling rate200kHz
Synchronous sourceU1~U6/I1~I6/EXT
Accuracy±(0.05% reading + 0.05% range)
Temperature drift±0.03% range/℃

Motor function –pulse frequency input parameters :

Input modeSafety BNC, floating, isolation, electrical isolation among A, B, Z of TORQUE and SPEED
Input impedance1MΩ±100kΩ
Frequency range1Hz~1MHz
Input amplitude range±22Vpeak
Maximum common mode voltage±42Vpeak
Effective amplitude1V
Minimum high pulse width2.5µS or more
Accuracy±(0.05% reading + 1mHz)

FFT computation function:

PA8000
Sampling rate/ time1k points5k points10k points50k points100k points200k points400k points500k points
2MS/s0.5ms2.5ms5ms25ms50ms100ms200ms250ms
1MS/s1ms5ms10ms50ms100ms200ms400ms500ms
500kS/s2ms10ms20ms100ms200ms400ms800ms1s
250kS/s4ms20ms40ms200ms400ms800ms1.6s2s
100kS/s10ms50ms100ms500ms1s2s4s5s
50kS/s20ms100ms200ms1s2s4s8s10s

Waveform sampling data storage function:

Storage itemsVoltage waveform, current waveform, computing waveform, FFT computation data, rotational speed, analog quantity of torque, harmonic data, custom function
Storage modesNormal, real-time, integration synchronization, conditional trigger
Data typeValue, waveform, value+waveform
File typeCSV format, PAD format
StorageInternal DOM, U disk

Integration function:

ModeThe manual, standard, continuous, real-time standard and real-time loop mode can be selected.
WP± modeCharging/discharging, electricity purchase/sale
TimerSet the timer to stop integration operations automatically. 0000h00m00s~10000h00m00s
Stop countingThe integration time and value will be maintained and the integration operation will be stopped when the integration time reaches its maximum value (10000 hours) or the integration value reaches the maximum/minimum display integration value (±999999M).
Accuracy±( power or current accuracy + time accuracy )
Time accuracy± 0.02% of reading

0.01% certified power measurement accuracy

PA8000 is a certified power analyzer with up to 0.01% power accuracy and 5MHz bandwidth. Used as the reference of energy efficiency tests for inverters, converters and power supply products, it is also the basis for laboratories to certify standard tests.

5MHz bandwidth for stable frequency characteristics and stability

PA8000 is the only power analyzer in the industry with up to 0.01% accuracy and up to 5MHz current and voltage frequency bandwidth, fully meeting the requirements of future power measurement of high-speed switching devices, such as sic.

High phase measurement accuracy under low power factors

PA8000 Power Analyzer has the bandwidth of DC/0.1Hz-5MHz. Thanks to the PLL source frequency-doubling technology, the harmonics measurement with faster speed and wider dynamic range has become realized. In harmonics mode, they can make accurate tests of fundamental voltage, fundamental current, power, phase, harmonic component and total harmonic distortion factor (THD).

1ms high data update rate

PA8000 is the only power analyzer in the world that data update rate can be set to 1ms.The 1ms~5ms high-speed data acquisition mode is added to PA8000. The 1ms data update rate can accurately capture the rapidly changing transient signals and perfectly display the waveform details. In motor applications, PA8000 is able to completely measure the startup waveform of the motor so as to provide important basis for users to assess the motor.

Phase compensation function of current sensor

When a current sensor is used as the input, the delay inherent to the sensor will increase the phase errors of the voltage and current. However, the phase compensation function of PA8000 can correct the phase error from the sensor and enhance the power measurement accuracy under the conditions of high frequency and low power factors so as to ensure the power measurement accuracy of the overall system.

Innovative Functions

  • PA8000 is equipped with the function of handwritten annotation on behalf of on-site annotation at any time and any place.

  • PA8000 adds a user-friendly graphic configuration interface to ensure testing and setup are completed once for all.

  • Users can also customize the graphical display in the instrument, with the testing items and numerical values in one-to-one match at a glance.

  • 12.1-inch HD display plus touch control to provide high-quality operating experience.